Study of Mechanical Characteristics of Electric Cupping Apparatus in Korea for Suggestion of its Assessment Guideline
Yi Seung-Ho

Kim Eun-Jung
Shin Kyung-Hoon
Nam Dong-Woo
Kang Jung-Won
Lee Seung-Deok
Lee Hye-Jung
Lee Jae-Dong
Kim Kap-Sung
Abstract
Objectives: The mechanical properties of Korean electric cupping systems are studied via experimental measurements. The study aimed at establishing the fundamentals of industrialization and systemization of oriental medicine device industry, as well as improving the quality of life for many Koreans.

Methods: We reviewed the studies on traditional cupping as well as modern one to fine necessary factors for electric cupping systems. To characterize the mechanical properties of Korean electric cupping systems, we measured the pressure characteristics of commercially available electric cupping system by using an automatic pressure acquisition system and a standard cup. The pumping capability was checked at 40 seconds, and the stability of the suction cup was checked at 600 seconds. We also acquired the noise level of each system in clinical setting. To check the portability of each system, we also measured its physical dimensions. We scrutinized system manuals provided by the system manufacturers.

Results: It took less than 5 second to reach the pressure if the connection between the air hose and the vacuum valve of the cupping system was secure. Pressure diminished to no more than 10% for 600s for all systems. Noise levels were 55¢¦70 dB. Increase in pressure was too fast to control for a designated vacuum level except for one product.
Conclusions: The Pumping ability of the systems is impressive and reliable. Pressure retention ability of each cup is quite reliable and reproducible. Therefore, their mechanical performances were worthy of recommendation. Some of them had noise level higher than 60 dB and they were bothersome. It was also suggested that the control for low to middle pressure needed to be accomplished by the cupping system.
KEYWORD
Electric cupping, vacuum pressure, noise level, manual