Journal of Soonchunhyang Medical College
1999 Volume.5 No. 1 p.99 ~ p.103
The Correspondence of Color Duplex Ultrasonography with Nocturnal Penile Tumescence Rigidity Test







Abstract
We intended to know that the result of color duplex ultrasonography corresponded to that nocturnal penile tumescence rigidity in the diagnosis of erctile dysfunction. We performed both color duplex ultrasonography and nocturnal penile tumescence rigidity in 15 erectile dysfunction men. Colr duplex ultrasonography was done with a 7 MHz. color Doppler unit after intracorporeal pharmacological injection and measured maximal arterail diameter, peak systolic velocity, end-diastolic velocity and the results were compared with rigidity and tumescence in nocturnal penile tumescence rigidity using RigiScan. The positive predictive value, negative predictive value, sensitivity, specificity, overall accuracy of color duplex ultrasonography was reasonable when the result of color duplex ultrasonography was compared with the that of nocturnal penile tumescence rigidity. According to this results, color duplex ultrasonography could not only predict but also correspond to nocturnal penile tumescence. Therefore, we suggest that color duplex ultrasonography can relpace nocturnal penile tumescence rigidity if psychogenic and neurogenic erectile dysfunction are distinguished by history and neurologic examination.