KMID : 0669519930030000059
|
|
Journal of Health Science - Yonsei University 1993 Volume.3 No. 0 p.59 ~ p.64
|
|
Design of a Volumetric Infusion Pump
|
|
|
|
Abstract
|
|
|
We designed a stable, accurate volumetric infusion pump by employing a peristaltic pump equipped with an anti-pulsation mechanism and driven by a microcomputer controlled stepping motor. This system is provided with the "Low battery", "Air-in-line", "Occlusion", "Bubble detection" and "Door open" alarms to ensure safe infusion. The error rate of this system is maximum ¡¾ 1.6%.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|