KMID : 0829320100130040157
Korean Journal of Clinical Microbiology
2010 Volume.13 No. 4 p.157 ~ p.161
Performance of the VITEK2 System for Detection of Inducible Clindamycin Resistance in Staphylococci
Kim Mi-Kyung

Hong Jong-Hee
Lee Mi-Ae
Abstract
Background: The Clinical and Laboratory Standards Institute (CLSI) recommends testing for inducible clindamycin resistance in clindamycin non-resistant and erythromycin resistant (CNR-ER) staphylococci by using a D-zone test. Recently, the VITEK2 system was developed to detect inducible clindamycin resistance in staphylococci. We evaluated the performance of the VITEK2 system by comparing it with a D-zone test.

Methods: In detecting inducible clindamycin resistance, a total of 142 clinical isolates of staphylococci were tested by using the VITEK2 Antimicrobial Susceptibility Test (AST)-P601 card (bioMerieux, Marcy l¡¯Etoile, France) and the D-zone test. Of the 142 isolates of staphylococci tested, 114 were CNR-ER staphylococci [40 coagulase-negative staphylococci (CoNS), 74 Staphylococcus aureus] and 28 were staphylococci, either resistant or susceptible to clindamycin and erythromycin (1 CoNS and 27 S. aureus).

Results: Of the 114 CNR-ER staphylococci, 98.6% (73/74) of S. aureus and 32.5% (13/40) of CoNS were inducible clindamycin resistant according to the Dzone test. Overall sensitivity and specificity of the VITEK2 system were 98.8% (85/86) and 98.2% (55/56) respectively, and the agreement between the VITEK2 system and the D-zone test was 98.6% (140/142).

Conclusion: The VITEK2 system shows high concordance with a D-zone test. The inducible clindamycin resistance in staphylococci can be detected easily and conveniently by the VITEK2 system.
KEYWORD
Inducible clindamycin resistance, Staphylococcus aureus, VITEK2