Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0858520080120010027
Journal of the Korean Society of Magnetic Resonance in Medicine
2008 Volume.12 No. 1 p.27 ~ p.32
Three-dimensional Chemical Shift Imaging with PRESS Excitation and Spiral Readouts
Kim Dong-Hyun

Abstract
Purpose: We developed a 3D CSI (chemical shift imaging) sequence that uses the PRESS (point resolved spectroscopy) excitation scheme and spiral-based readout gradients.

Material and Methods :We implemented constant-density spirals ( matrix, FOV) which use analytic equations to enable real-time prescription on the scanner. In-vivo data from the brain were collected and reconstructed using the gridding algorithm.

Results:Data illustrate that with our imaging sequence, the benefits of the PRESS technique, which include elimination of lipid artifacts, remain intact while flexible scan time versus resolution tradeoffs can be achieved using the constant-density spirals. Volumetric high resolution 3D CSI covering 5760 cm3 could be obtained in 12.5 minutes.

Conclusion:Spiral-based readout gradients offer a flexible tradeoff between scan time versus resolution. By combining this feature with PRESS based excitation, efficient methods of volumetric spectroscopic imaging can be accomplished by obtaining whole brain coverage while eliminating lipid contamination.
KEYWORD
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø