Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0858520090130020152
Journal of the Korean Society of Magnetic Resonance in Medicine
2009 Volume.13 No. 2 p.152 ~ p.160
Contrast-Enhanced Magnetic Resonance Angiography for Evaluation of the Steno-occlusive Disease of the Supraaortic Arteries: Comparison with Computed Tomography Angiography and Digital Subtraction Angiography
Jeh Su-kyoung

Lee Kwang-Soo
Kim Yeong-In
Kim Bum-Soo
Ahn Kook-Jin
Jung So-Lyung
Shin Yong-Sam
Lee Kwan-Sung
Abstract
Purpose : To intra-individually compare diagnostic accuracy of high-resolution contrast-enhanced magnetic resonance angiography (CE-MRA) with computed tomography angiography (CTA) and digital subtraction angiography (DSA) for the assessment of supraaortic steno-occlusive disease.

Materials and Methods : Twenty-eight patients (20 men, 8 women, 53-79 years of age) underwent supraaortic CE-MRA, CTA and DSA. CE-MRA was performed on two 1.5T MR scanners (voxel dimension: or ), and CTA on 64-slice CT scanners (voxel dimension: ). All the three examinations were completed within 40 days (median 19 days; range 1-40 days). Retrospective evaluation and measurement of diameter of 6 extracranial and 9 intracranial arterial segments was done by 2 experienced radiologists.

Results: A total of 420 arterial segments were examined by CE-MRA, CTA and DSA. On DSA, 34 stenoocclusive lesions were noted at extracranial (n= 19) and intracranial (n = 15) vessels. For extracranial stenosis greater than 70%, sensitivity, specificity, positive predictive value (PPV) and negative predictive values (NPV) were 94.7%, 98.7%, 90.0% and 99.3% on CE-MRA, and 94.7%, 99.3%, 94.7% and 99.3% on CTA. For intracranial stenosis greater than 50%, sensitivity; specificity, PPV and NPV were 93.3%, 98.3%, 77.8%and 99.6% on CE-MRA, and 86.7%, 97.9%, 72.2% and 99.1 % on CTA, with DSA as the standard of reference.

Conclusion : Supraaortic CE-MRA is as reliable as CTA in depicting the arterial stenosis, and is effective in screening of significant stenosis of both extracranial and intracranial arterial stenosis.
KEYWORD
Diagnostic accuracy, Sensitivity and specificity, Contrast enhanced MR angiography
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø