KMID : 0861419990040010164
|
|
Korean Journal of Nuclear Medicine Technology 1999 Volume.4 No. 1 p.164 ~ p.169
|
|
|
|
ÀÌ¿µÈñ
¿À½ÂÁØ/À¯ÈñÀç/±èÈ£¼º/½Å»ó±â/Á¶½Ã¸¸/Á¤À±¿µ
|
|
Abstract
|
|
|
Purpose: We evaluated a rapid preparation procedures for the labeling and quality control of
99mTc-ECD, MIBI and MAG3 using microwave heating and Sep-Pak cartridges.
Materials and Methods: 99mTc labeling of ECD, MIBI and MAG3 kit
preparation was performed according to the package inserts with microwave heating modification.
Heating time was 10¡15sec, and heating was performed with 3mm plastic bottle with screw cap to
prevent radiation contamination. Labeling efficiency was obtained with C18 or Alumina
N Sep-Pak cartridge.
Results: The radiochemical purity of 93¡96% for 99mTc-ECD and 95¡99% for
99mTc-MIBI was obtained using Alumina N Sep-Pak cartridge. The optimum irradiation
time of microwave method for 3ml 99mTc-labeled radiopharmaceutical solution was 10
sec for 99mTc-ECD, MIBI and 15 sec for 99mTc-MAG3. The
Alumina N Sep-PAk cartridge was used for checking of labeling efficiency of 99mTc-ECD,
and 99mTc-MIBI. The results of quality control data with
Sep-Pak cartridge were well correlation with TLC method data. The total preparation time
of these radiopharmaceutical was 5¡6min including quality control procedure.
Conclusion: This study demonstrates that radiopharmaceuticals preparation by microwave
heating and quality control by Sep-Pak cartridges can be efficiently utilized as an alternative
to the recommended method by manufacturer's manual.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|