Korean Journal of Nuclear Medicine Technology
2010 Volume.14 No. 1 p.115 ~ p.121
The Effect of Using Two Different Type of Dose Calibrators on In Vivo Standard Uptake Value of FDG PET
Park Young-Jae

Bang Seong-Ae
Lee Seung-Min
Kim Sang-Eon
Lee Kyung-Jae
Lee In-Won
Abstract
Purpose:The purpose of this study is to measure F-18 FDG with two different types of dose calibrator measuring radionuclide and radioactivity and investigate the effect of F-18 FDG on SUV (Standard Uptake Value) in human body.

Materials and Methods: Two different dose calibrators used in this study are CRC-15 Dual PET (Capintec) and CRC-15R (Capintec). Inject 1 mL, 2 mL, 3 mL of F-18 FDG into three 2 mL syringes, respectively, and measure initial radioactivity from each dose calibrator. Then measure and record radioactivity at 30 minute interval for 270 minutes. According to the initial radioactivity, linearity between decay factor driven from radioactive decay formula and the values measured by dose calibrator have been analyzed by simple linear regression. Fine linear regression line optimizing values measured with CRC-15 through regression analysis on the basis of the volume of which the measured value is close to the most ideal one in CRC-15 Dual PET. Create ROI on lung, liver, and region part of 50 persons who has taken PET/CT test, applying values from linear regression equation, and find SUV. We have also performed paired t-test to examine statistically significant difference in the radioactivity measured with CRC-15 Dual PET, CRC-15R and its SUV.

Results:Regression analysis of radioactivity measured with CRC-15 Dual PET and CRC-15R shows results as follows: in the case 1 mL, the r statistic representing correlation was 0.9999 and linear regression equation was y=1.0345x+0.2601; in 2 mL case, r=0.9999, linear regression equation y=1.0226x+0.1669; in 3 mL case, r=0.9999, linear regression equation y=1.0094x+0.1577. Based on the linear regression equation from each volume, t-test results show significant difference in SUV of ROI in lung, liver, region part in all three case. P-values in each case are as follows: in 1 mL case, lung, liver and region (p<0.0001); in 2 mL case, lung (p<0.002), liver and region (p<0.0001); in 3 mL case, lung (p<0.044), liver and region (p<0.0001).

Conclusion:Radioactivity measured with CRC-15 Dual PET, CRC-15R, dose calibrator for F-18 FDG test, do not show difference correlation, while these values infer that SUV has significant differences in the aspect of uptake in human body. Therefore, it is necessary to consider the difference of SUV in human body when using these dose calibrator.
KEYWORD
Dose calibrator, SUV, Linear regression equation, Regression analysis, Paired t-test