Usefulness of Xact-bone for the Resolution Advancement of Gamma Camera Image
Kim Jong-Pil

Yoon Seok-Hwan
Lim Jung-Jin
Woo Jae-Ryong
Abstract
Purpose:The Boramae Hospital are currently using Wide beam reconstruction (WBR: UltraSPECT, Israel) to improve the resolution. The Xact-bone belongs to the WBR. It has been reported that Xact-bone helps us to improve image resolution and contrast. This study will be evaluated clinical usefulness of Xact-bone method.

Materials and Methods: The usefulness evaluation of Xact-bone method was analyzed in resolution test and contrast ratio. The resolution test in Planar image were obtained from Full width at half maximum (FWHM) by using capillary tube. And the contrast ratio was obtained from Bone and Soft tissue (B/S) ratio values that were acquired from bone scan study of 50 patients before and after using the Xact-bone method. We prepared the Triple Line Source Phantom, NEMA IEC Body Phantom and Standard Jaszczak Phantom to acquire the FWHM and Contrast Ratio values of Single photon emission computed tomography (SPECT) image. Subsequently we compared among the Filtered backprojection (FBP), Orderd subset expectation maximization (OSEM) and Xact-Bone image.

Results:The results of the planar Xact-bone data improved resolution about 20% by using capillary tube. In addition it was improved B/S ratio about 15%. When using Triple Line Source Phantom, SPECT Xact-bone data improved resolution for both FBP, OSEM methods about 20% and 10%, respectively. Contrast ratio in each spheres has also been increased for both methods that using NEMA IEC body Phantom and Standard Jaszczak Phantom.

Conclusion:When we were using Xact-bone method, we could see to improve the resolution and Contrast ratio as compared to do not use the Xact-bone method. Accordingly, by using WBR¡¯s Xact-bone method is expected to improve the image quality. However, when introducing new software, it is needed to match the characteristics of the hospital protocol and clinical application.
KEYWORD
Xact-Bone, FBP, OSEM, Resolution, Contrast ratio