KMID : 0861420110150020083
Korean Journal of Nuclear Medicine Technology
2011 Volume.15 No. 2 p.83 ~ p.87
The Evaluation of TrueX Reconstruction Method in Low Dose
Oh Se-Moon

Kim Kye-Hwan
Kim Seung-Jeong
Lee Hong-Jae
Kim Jin-Eui
Abstract
Purpose:Recently in diagnostics area PET/CT is using a variety of areas including oncology, as well as in cardiology, neurology, etc. While increasing in the importance of PET/CT, there are various researches in the image quality related to reconstruction method. We compared and tested Iterative 2D Reconstruction Method with True X Reconstruction method by Siemens through phantom experiment, so we can see increasing of clinical usefulness of PET/CT. Materials and

Methods: We measured contrast ratio and FWHM due to evaluating images on dose and experiment using Biograph 40 True Point PET/CT (Siemens, Germany). Getting a result of contrast ratio and FWHM, we used NEMA IEC PET body phantom (Data Spectrum Corp.) and capillary tube. We used the current TrueX and the previous Iterative 2D algorithm for all images which have 10 minutes long. Also, a clinical suitability of parameter for Iterative 2D and a recommended parameter by Siemens for True X are applied to the experiment.

Results:We tested FWHM using capillary tube. As a result, TrueX was less than Iterative 2D. Also, the differences of FWHM get bigger in low dose. On the other hand, we tested contrasts ratio using NEMA IEC PET body phantom. As a result, TrueX was better aspect than Iterative 2D. However, there was no difference in dose.

Conclusion: In this experiment, TrueX get higher results of contrast ratio and spatial resolution than Itertive 2D through experiment. Also, in the reconstruction result through TrueX, TrueX had better aspect of resolution than Iterative 2D in low dose. However, contrast ratio had no specific difference. In other words, TrueX reconstruction method in PET/CT had higher clinical value in use because TrueX can reduce exposure of patient and had a better quality of screen.
KEYWORD
Iterative 2D, TrueX, spatial resolution, FWHM, PET/CT