KMID : 0861420120160010070
Korean Journal of Nuclear Medicine Technology
2012 Volume.16 No. 1 p.70 ~ p.75
Comparison of Image Uniformity Due to Position Shifting in COR on Myocardial SPECT
Lim Hyun-Jin

Kim Joong-Hyun
Kim Jae-Il
Lim Jung-Jin
Kim Jin-Eui
Kim Hyun-Joo
Lee Jae-Sung
Lee Dong-Soo
Abstract
Purpose: It is important to acquire accurate data because the SPECT scan affected by various physical factors. The aim of this study was to compare the uniformity when both centers were matched or mismatched differed from position of heart in COR.

Materials and Methods: The images were acquired with cylindrical uniform phantom (6.7 cm diameter, 9 cm length) and heart insert phantom using Cardio MD SPECT system (Philips, USA). The phantoms were positioned on COR as well as four different points which were 10 cm above, below, left and right side from the COR. The counts from the both edge of cylindrical uniform phantom and those from the both wall of heart insert phantom were compared by using vertical and horizontal line profile. In addition, the qualitative evaluation was performed with heart insert phantom images and volunteer test.

Results: In heart insert phantom study, the differences of counts between COR and 10 cm above, below, left and right point of COR were 1.1, 4.1, 4.9, 2.2 and 0.9% using T-A curve for horizontal view. In case of vertical view of COR 3.9, 21.9, 3.5, 23.9, 14.0% were shown. In cylindrical phantom study, the differences of counts between COR and 10 cm above, below, left and right point of COR were 4.3, 0.3, 3.3, 2.6 and 0.7% using T-A curve for horizontal view. In case of vertical view of COR 2.7, 3.0, 1.0, 0.3, 3.4% were shown. For qualitative evaluation, the images at COR were the most uniform for both of heart insert phantom and volunteer test, whereas other four positions showed somewhat distorted images.

Conclusion: It showed the most uniform images when COR is matched with the heart. Therefore, we can expect that distortion which increased or decreased of myocardial perfusion will be prevented by matching the heart and COR when positioning. Furthermore, the accuracy of diagnosis will be improved as well.
KEYWORD
COR matching, Uniformity