Korean Journal of Nuclear Medicine Technology
2012 Volume.16 No. 1 p.91 ~ p.95
The Study on Ejection Fraction Change According to Patient Position Difference in Gated Blood Pool Scan
Choi Ho-Yong

Kim Da-Eun
Jeong Jae-Hoon
Yun Sang-Hyeok
Kim Yeong-Seok
Won Woo-Jae
Abstract
Purpose: In this study, we evaluated the ejection fraction (EF) according to the difference of patient position in Gated Blood Pool (GBP) scan.

Material and Methods: The analysis was performed to 80 patients (51.2¡¾17.4 years old) who examined GBP scan in Department of nuclear medicine, National Cancer Center from March 2011 to August 2011. We divided the patients into two groups; one group received conventional position (raise left arm up supine) and supine position (group 1) and the other group received conventional position and left arm back down supine position (group 2). To observe the change EF according to patient position difference, the image was reconstructed and analyzed by Xeleris (GE, USA). We measured body mass index (BMI) of patients.

Result: In group 1, EF error less than 3% occurred at a rate of 72.5% (29 of the 40 patients). In group 2, EF error less than 3% occurred at a rate of 79% (32 of the 40 patients). The patient¡¯s BMI did not affect ejection fraction.

Conclusion: The EF error of left arm back down supine position closer to conventional position than in supine position shows the results.
KEYWORD
Gated Blood Pool (GBP), Ejection Fraction (EF), Patient Position