Radioimmunoassay Using Quality Control Materials BIO-RAD of the Suggestions for Setting Acceptable Range
Park Hee-Won

Kim Tae-Hoon
Noh Kyeong-Woon
Kim Hyun-Joo
Abstract
Purpose: Immunosorbent assay most commonly used in the laboratory and commercial third-party quality control material is a substance that provided by BIO-RAD. However, in Reference Sheet by radioimmunoassay test kit or a measuring device for the mention of Acceptable Range is somewhat lacking. Radioimmunoassay for the inspection of test results by setting Acceptable Range to increase the objectivity of the recommendations on the data accumulated by the manufacturer listed in the Reference Sheet is to be issued.

Materials and Methods: In our hospital since 2009 partially BIO-RAD using quality control materials in 2011, excluding certain items, some items were most of the BIO-RAD third-party quality control materials are used. Thus, internal quality control data accumulated BIO-RAD¡¯s Unity Real Time program by using the items were measured.

Results: BIO-RAD using quality control material items were about 50 of the 20 Point Data averages, standard deviations, variation coefficients were calculated to measure the Acceptable Range of kit, automated immunoassay attributed Roche Elecsys / E170 / cobas e Systems the measures and compared.

Conclusions: BIO-RAD QC materials commonly used hospital and peer group by setting the measurement kit, suitable for laboratory equipment for radioimmunoassay of Acceptable Range manufacturer recommendation to increase the objectivity of the test results by national and international recognition for radioimmunoassay should seek to increase.
KEYWORD
Quality control materials, Acceptable Range