KMID : 0895419970070020223
Journal of Korean Society of Occupational and Enviromental Hygiene
1997 Volume.7 No. 2 p.223 ~ p.232
Accuracy and Precision of Ion Chromatography / Visible Absorbance Detection for Analyzing Hexavalent Chromium Collected on PVC Filter



Abstract
The accuracy and precision of a modified method of NIOSH Method 7600 and EPA method 218.6 was determined for analyzing hexavalent chromium, Cr(¥µ), collected on PVC filter from workplace air. The method was designed to extract from Cr(¥µ) on PVC filter with a alkali solution, 2% NaOH/3% Na©üCO©ý, and to analyze it using ion chromatography/visible absorbance detection (IC/VAD). The results and conclusion are as the following.
1. The peak of Cr(¥µ) was separated sharply on chromatogram and was linearly related with Cr(¥µ) concentration in solution. The correlation coefficient was 0.9999 in a calibration curve. The limit of detection was 0.25 §¶/sample.
2. The accuracy(% recovery) was 93.3% in a set of samples(9-50 §¶) stored for a day, and 100.1 %(l0-60 §¶) in another set of samples stored for 2 hours. It is assumed that the difference in recovery by storage time was due to reduction of Cr(¥µ) to (Cr(¥³).
3. The precision(coefficient of variation, CV) of the method was 0.015 in spiked samples with Cr(¥µ) standard solution, and 0.010 in spiked samples with plating solution from a chrome electroplating factory. The overall CV in ail types of samples was 0.013.
4. The Cr(¥µ) was stable in 2% NaOH/3% Na©üCO©ý at least for 10 hours.
In conclusion, the IC/VAD method is appropriate for determining low-level Cr(¥µ) in workplace air containing various interferences.