KMID : 0895419980080010124
|
|
Journal of Korean Society of Occupational and Enviromental Hygiene 1998 Volume.8 No. 1 p.124 ~ p.132
|
|
Field Evaluation of Korean Passive Sampler for Organic Vapor
|
|
|
|
Abstract
|
|
|
The Korean-made passive samplers were evaluated at the working environment for field testing. Tested materials were n-hexane, toluene and trichloroethylene.
The performance of passive samplers depended on types and concentrations of organic vapors. Sampling rates were not steady until certain concentrations. The optimum concentration for determination of airborne toluene by passive samplers was equal to or over 10 ppm which is 1/10 of the Korean occupational exposure limit. Optimum concentration of n-hexane was equal to and over 1 ppm which is 1/50 of Korean occupational exposure limit. But for trichloroehtylene, coefficient of variation was 53.5 %. Passive samplers may be used for determination of n-hexane. For other materials, further study on the performance of Korean-made passive samplers is required.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|