Journal of Korean Society of Occupational and Enviromental Hygiene
2003 Volume.13 No. 2 p.113 ~ p.118
Reduction Rate of Hexavalent Chromium Mist in Air and on PVC Filter during Air Sampling in Field Plating Process




Abstract
Hexavalent chromium, Cr(¥µ),may reduce on polyvinyl chloride (PVC) filter during air sampling. This study was conducted to cstimate the reduction rate of Cr(¥µ) on PVC filter during air sampling and an correction equation for th loss of Cr(¥µ) by feduction in an industrial plating process. The reduction of Cr(¥µ) was progressed rapidly in 60 minutes and slowly thereafter. The empirical reduction rate of Cr(¥µ) (Y) over sampling time (X, min) was the following : Y= 0.916 -0.0208¡îx (r=0.73, p¡´0.05). The loss of Cr(¥µ) caused by reduction during sampling can be corrected using the following empirical equation : Sc = S x 1/Y. where Sc = corrected value, S = analytical result, and Y= reduction rate of Cr(¥µ).