Development of an Evaluation Method for the Flow Rate Performance of Gas Sampling Pumps Using Adsorbent Tubes
Kim Nam-Hee

Song Ho-June
Kim Ki-Youn
Ma Hye-Lan
Yi Gwang-Yong
Jeong Jee-Yeon
Abstract
Objectives: Flow rate stability is very important for obtaining reliable measurements. However there is no easily used method for checking whether the flow rate set at the initial stage is sustained during sampling. The purpose of this study was to develop a method to evaluate the flow rate performance of gas sampling pumps with adsorbents commonly used to sample gases.

Materials and methods: We tested the back pressure of gas sampling pumps commonly used in Korea with adsorbents such as charcoal and silica tubes and attempted to discover the combination conditions of adsorbents in accordance with back pressure required by ISO standard 13137.

Results: We found the combination of sampling adsorbents to be applicable to the pressure drop required by the ISO standard for evaluating flow rate stability under increasing pressure drop and long term (eight-hour) performance.

Conclusions: This evaluation method of using a sampling media matrix for checking flow rate stability as proposed by this study could be a highly useful tool for determining the reliability of the performance pumps before sampling.
KEYWORD
Adsorbents, back pressure, flow rate, pump, ISO