Korean Journal of Oral and Maxillofacial Radiology
2007 Volume.37 No. 3 p.149 ~ p.156
Absorbed and effective dose from periapical radiography by portable intraoral x-ray machine
Choi Jeong-Yeon

Han Won-Jeong
Kim Eun-Kyoung
Abstract
Purpose : The purpose of this study was to measure the absorbed dose and to calculate the effective dose for periapical radiography done by portable intraoral x-ray machines.

Materials and Methods : 14 full mouth, upper posterior and lower posterior periapical radiographs were taken by wall-type 1 and portable type 3 intraoral x-ray machines. Thermoluminescent dosemeters were placed at 23 sites at the layers of the tissue-equivalent ART woman phantom for dosimetry. Average tissue absorbed dose and radiation weighted dose were calculated for each major anatomical site. Effective dose was calculated using 2005 ICRP tissue weighted factors.

Results : On 14 full mouth periapical radiographs, the effective dose for wall-type x-ray machine was 30 Sv; for portable x-ray machines were 30 Sv, 22 Sv, 36 Sv. On upper posterior radiograph, the effective dose for wall-type x-ray machine was 4 Sv; for portable x-ray machines doses were 4 Sv, 3 Sv, 5 Sv. On lower posterior radiograph, the effective dose for wall type x-ray machine was 5 Sv; for portable x-ray machines doses were 4 Sv, 4 Sv, 5 Sv.

Conclusion : Effective doses for periapical radiographs performed by portable intraoral x-ray machines were similar to doses for periapical radiographs taken by wall type intraoral x-ray machines. (Korean J Oral Maxillofac Radiol 2007; 37 : 149-56)
KEYWORD
Radiometry, Radiation Monitoring, Thermoluminescent Dosimetry, Diagnostic Imaging, Radiography, Dental