Korean Journal of Oral and Maxillofacial Radiology
2009 Volume.39 No. 4 p.199 ~ p.203
Reference dose levels for dental panoramic radiography in Anyang City
Kim Young-Hee

Lee Jae-Seo
Yoon Suk-Ja
Kang Byung-Cheol
Abstract
Purpose : To measure dose-width product (DWP) values used for dental panoramic radiography in Anyang city, Korea.

Materials and Methods : Thirty-six panoramic dental radiographic sets (17 analogue panoramic sets and 19 digital panoramic sets) in 36 dental clinics in Anyang city were included in the study. Each patient¡¯s panoramic exposure parameters were simulated and the panoramic radiation doses were measured at the secondary collimator using a Mult-O-Meter (Unfors Instruments, Billdal, Sweden) at each dental clinic during 2006. The third quartile DWP was determined from 310 surface dose measurements on adult.

Results : The third quartile DWP for adult panoramic radiograph was 106.7 mGy mm. For analogue and digital panoramic radiograph, 3/4 DWP were 116.8 mGy mm and 72 mGy mm respectively. The overall third quartile DWP of panoramic radiography was 106.7 mGy mm.

Conclusion : The measured 3/4 DWPs were higher than the 3/4 DWP of 65 mGy mm recommended by NRPB. Dentists who are operating above the reference dose should lower their panoramic exposure doses below the recommended reference value by changing the exposure parameters and/or their panoramic equipments.
KEYWORD
Radiation, Gray, Panoramic Radiography