KMID : 0896219940140011043
Journal of Daegu Health College
1994 Volume.14 No. 1 p.1043 ~ p.1064
The Compensation of Energy Dependence of TLD Sensor and Its Characteristios
Park Myeong-Hwan

Abstract
The CaSO©þ:Tm-PTFE TLDs have been prepared and their energy dependences of TL intensity to X-rays have been investigated. The dose dependence of the prepared CaSO©þ:Tm-PTFE TLDs is linear within the range of 10uGy-10Gy for X-rays and gamma rays. The spectral peaks of TL emission spectrum are at about 350§® 475§®. The fading rate of the main peak has been found to be about 2% for 30days when CaSO©þ:Tm-PTFE TLDs stored in the dark room at room temperature. The energy dependence of CaSO©þ:Tm-PTFE TLD has been compensated by combining the TL responses of one bare TLD and five TLDs filtered with 10.§® Al, 0.2§® Cu, 0.5§® Cu, 1.5§® Cu, 2.0§® Pb respectively. The determined correction coefficients for filter combination are a_(0)=0.078, a©û=-0.009, a©ü=-0.235, a©ý=0.588, a©þ=0.568 and a_(5)=0.180 respectively. From the results of these studies, the prepared TLD badge of six dosimeter combination may be useful as a radiation dosimeter for personal and environmental monitoring.
KEYWORD