Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0896219990190010055
Journal of Daegu Health College
1999 Volume.19 No. 1 p.55 ~ p.61
Determination of Dose Distribution under Nonequivalent Condition by Prosthesis in RT
Lee Joon-Il

Kim Seong-Hwan
Abstract
When a patient was irradiated with prosthetic hip, the dose distribution was changed according to inhomogeneous materials. The density, effective atomic number, and the composition of material had influence on absorbed dose distribution. In this study, the influence of inhomogeneous material(Ti) was measured using a polyethylene phantom, which consisted of various diameter of titanium, with film dosimetry. As a result, the backward dose showed 29.5% increase by backscattering, the forward dose showed 28% decrease by absorption, and the side dose showed 7% increase by scattering, when ¡Ë25 §® diameter Ti was used. In addition forward dose was in inverse proportion to the thickness of prosthetic material. When the prosthetic hip of patient is an irradiated filed, we must carefully study the absorbed dose distribution.
KEYWORD
FullTexts / Linksout information
Listed journal information