In this study, visual recognition of X-ray images were estimated using hawlet charts. The resolution has achieved excellent result when densities of film were 1.0~1.5. The more thickness of object was increase, the more the resolution was decrease, because scattered X-ray were increased. The more distance of object to film was increase, the more the resolution was decrease, because groedel effect. The more focal spot size was increase, the more resolution was decrease, because unsharpness was increase. The more field size was increase and grid ratio was decrease, the more resolution was decrease, because scattered X-ray were increase. As a result, we could to improve image quality by determination of resolution according to Radiography parameters. And, we proved that hawlet charts were useful for estimate of image quality and set up x-ray radiography.
|