Spontaneous Fracture and Embolization of Implantable Venous Access Device in Breast Cancer Patient
Kim Young-Suk

Kim Eun-Young
Park Woo-Chan
Abstract
Totally implantable venous access devices (chemoports) are used increasingly in oncology patients to provide simple and durable access for intravenous delivery of drugs and fluids. However, these catheters are associated with a number of complications, such as infection, thrombosis, and stenosis. Fracture of a venous catheter is also a rare complication, known as "pinch-off syndrome." A 43-year-old female with breast cancer was presented with chest pain of one week¡¯s duration after undergoing placement of a chemoport for adjuvant chemotherapy. Her chest X-ray showed that a portion of the catheter had broken free and was lodged in the right ventricle. The embolic fragment was successfully retrieved via percutaneous femoral vein insertion of a snare.
KEYWORD
Breast neoplasms, Catheters, Fracture