High Performance Liquid Chromatographic Assay of Ofloxacin in Plasma


Abstract
A high-performance liquid chromatographic method with fluorometric detection was evaluated for analysis of ofloxacin in plasma. Biological fluids (plasma, 200;{mu}L) were prepared for assay by protein precipitation with chlorofurm. The detection of ofloxacin and triamterene as an internal standard were performed at 358 nm for excitation and 495 nm for emission. The HPLC separation was carried out on Ultrasphere ODS column (4.6 mm{times}25;cm,;5;{mu} m) with acetonitrile (45%)-phosphoric acid $(1.5%);containing;0.3% sodium laurylsulfate as the mobile phase. The flow-rate was 1.0 mL/min. The calibration graphs were linear from 3.0 to 80 ng/mL with r=0.998. The minimal detectable concentration in plasma was 1.5 ng/mL. The proposed technique is reproducible, selective, reliable and sensitive.