Korean Journal of Clinical Pharmacy
2009 Volume.19 No. 1 p.65 ~ p.68
Determination of Flow Rate and Stability of 5-Fluorouracil in Disposal Infusion Device, Anapa
Kim Jung-Tae

Chung Sung-Hyun
Abstract
Disposal infusion device is known to be useful for chemotherapy. Anti-cancer drug can be released by the force of carbon dioxide or balloon. In this study, we compared the Anapa¢ç (LC0020) with B Company (LV2 ml) in terms of infusion rate and stability. Infusion rate was determined every six minute using software, MSI08IH. Stability of 5-fluorouracil was examined periodically using a High Performance Liquid Chromatography. Infusion rates of gas-derived Anapa device were 2.29, 1.86, 1.98 ml/hr and those of balloon-derived B Company device were 1.71, 1.58, 1.37 ml/ min. There were no significant differences in stability of 5-fluorouracil between Anapa¢ç and B Company devices. In summary, gas-derived Anapa¢ç device is thought to be comparable or superior to balloon-derived B Company device as far as infusion rate and stability are concerned. We expect that Anapa¢ç as a home infusion device can be employed to improve a quality of life and compliance of cancer patients.
KEYWORD
Infusion, Anapa¢ç, Flow rate, Stability, 5-fluorouracil