Journal of Health Science & Medical Technology
2004 Volume.30 No. 2 p.31 ~ p.37
A experimental study on conversion X-ray exposure factors according to plaster thickness and synthetic cast
Kim Jung-Min

Jung Hea-Won
Abstract
Composition and X-ray transmission Property very different plaster of paris to synthetic cast. We experiment on conversion exposure factor according to thickness of plaster of paris and synthetic cast. From the result of the this study, it could be concluded as follows. In synthetic cast radiography(3 layer, 4 §®) can make same density with changes in 8 kVp or 53%mAs. Wet plaster of paris(6 layer, 5.3 §®) can make same density with change of up to 11 kVp or 81%mAs. To the contrary dry plaster of paris(6 layer, 5.3 §®) can make same density with change of up to 10kVp or 73%mAs.
KEYWORD
plaster of paris, synthetic cast, exposure correction