Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0980320170170020135
Journal of Dental Anesthesia and Pain Medicine
2017 Volume.17 No. 2 p.135 ~ p.138
Airway obstruction by dissection of the inner layer of a reinforced endotracheal tube in a patient with Ludwig's angina: A case report
Shim Sung-Min

Park Jae-Ho
Hyun Dong-Min
Lee Hwa-Mi
Abstract
Intraoperative airway obstruction is perplexing to anesthesiologists because the patient may fall into danger rapidly. A 74-year-old woman underwent an emergency incision and drainage for a deep neck infection of dental origin. She was orally intubated with a 6. 0 mm internal diameter reinforced endotracheal tube by video laryngoscope using volatile induction and maintenance anesthesia (VIMA) with sevoflurane, fentanyl (100 ¥ìg), and succinylcholine (75 mg). During surgery, peak inspiratory pressure increased from 22 to 38 cmH2O and plateau pressure increased from 20 to 28 cmH2O. We maintained anesthesia because we were unable to access the airway, which was covered with surgical drapes, and tidal volume was delivered. At the end of surgery, we found a longitudinal fold inside the tube with a fiberoptic bronchoscope. The patient was reintubated with another tube and ventilation immediately improved. We recognized that the tube was obstructed due to dissection of the inner layer.
KEYWORD
Intubation, Bronchoscopes, Ludwig's Angina, Reinforced Endotracheal Tube
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)