KMID : 0981820070270030162
Korean Journal of Laboratory Medicine
2007 Volume.27 No. 3 p.162 ~ p.168
Evaluation of the Abbott Cell-Dyn Sapphire Hematology Analyzer
Park Youn-Hee

Song Jae-Woo
Song Sung-Wook
Song Kyung-Soon
Ahn Mee-Suk
Yang Mi-Suk
Kim Il
Choi Jong-Rak
Abstract
Background : The performance of Cell-Dyn Sapphire (Abbott Diagnostic, USA) was compared to the Bayer Advia 2120 (Bayer Diagnostics, USA), Sysmex XE-2100 (Sysmex Corporation, Japan), and reference microscopy.

Methods : Three hundred samples for routine CBC and WBC differentials were randomly chosen for a comparison analysis. The Cell-Dyn Sapphire system was evaluated according to the linearity, imprecision, inter-instrument correlations, and white blood cell differential.

Results : The CBC parameters (WBC, RBC, hemoglobin and platelet) showed a significant linearity with correlation coefficients greater than 0.99 (P<0.0001). Coefficients of variation (CV) for withinrun and differential count of WBC were less than 5% except for Total CV for monocytes, eosinophils, and basophils and within-run CV for low valued eosinophils. The correlation coefficients with manual count were lower in monocytes, eosinophils, and basophils than in neutrophils and lymphocytes. The correlation with other hematology anlayzers was significant exclusive of basophils.

Conclusions : These results demonstrate that the Cell-Dyn Sapphire has a good linearity, an acceptable reproducibility, a minimal carryover, and a comparable performance with the sysmex XE-2100 and Advia 2120. (Korean J Lab Med 2007;27:162-8)
KEYWORD
Cell-Dyn Sapphire, Performance, Correlation