KMID : 0981820070270030210
Korean Journal of Laboratory Medicine
2007 Volume.27 No. 3 p.210 ~ p.215
Evaluation of the OC-SENSOR neo System for Testing Fecal Occult Blood
Kim Ji-Hyun

Chung Hee-Jung
Yoon Nam-Seob
Pyo Yeon-Jeong
Kim Mi-Na
Bae Hye-Kyung
Abstract
Background : Fecal occult blood tests (FOBTs) have been widely used as a means of colorectal cancer screening. Automated FOBTs using immunologic principles have the advantages such as quantitation, high specificity, and high throughput. We evaluated a newly-introduced automated FOBT analyzer, OC-SENSOR neo (OC neo) (Eiken Chemical Co., Japan).

Methods : The precision, linearity, and carry-over rate of OC neo were assessed with specimens prepared in accordance with the guidelines of CLSI. We performed a parallel test between OC neo and OC-SENSOR I (OC I) (Eiken Chemical Co.) using 300 consecutive stool specimens and 60 OC I-positive specimens. The results were analyzed with SPSS version 13.0 (SPSS Inc., USA).

Results : The coefficients of variation (CV) of within-run, between-run, and between-day using OCControl L (Eiken Chemical Co.) of ca. 150 ng/mL were 3.5-7.8%, 4.5-8.8% and 4.9-5.0%, respectively. The linear regression coefficient and carry-over rate with the range of 67.8-939.4 ng/mL were 0.9998 (P<0.001), and 0.1%, respectively. Correlation coefficient between OC neo and OC I was R2=0.954 (P<0.001) for 60 OC I-positive specimens. The positive and negative interpretations of 300 consecutive specimens by OC neo were completely consistent with those of OC I.

Conclusions : Because OC neo showed an excellent performance and a good correlation with OC I, OC neo warrants to be a reliable quantitative FOBT system for high volume laboratories. (Korean J Lab Med 2007;27:210-5)
KEYWORD
Fecal occult blood test, Colorectal cancer, OC-SENSOR neo, Automation