Korean Journal of Laboratory Medicine
2009 Volume.29 No. 4 p.324 ~ p.330
The Optimal Wavelength of Spectrophotometric Broth Microdilution Antifungal Susceptibility Test for Flucytosine and Three Azoles
Lee Ja-Young

Chang Chul-Hun L.
Kho Weon-Gyu
Lee Eun-Jung
Shin Jeong-Hwan
Kim Hye-Ran
Kim Si-Hyun
Jeong Heang-Soon
Oh Seung-Hwan
Lee Jeong-Nyeo
Abstract
Background: There is no guideline for the appropriate wavelength at which to measure the optical density (OD) value in broth microdilution antifungal susceptibility testing, although a spectrophotometric reading method is commonly used. The present study aimed to analyze the difference in the OD values over the range of visible light and to ascertain the optimal wavelength for the spectrophotometric method of microdilution testing.

Methods: We measured the OD of background blank controls of broth medium, antifungal agents, and inocula of five type strains using a Synergy HT multi-detection microplate reader at 5-nm intervals from 380 nm to 760 nm. We also estimated the OD differences between the 50% of growth control and blank control.

Results: The OD of the blank control showed a parabola shape with two peaks and steadily decreased at longer wavelengths. The curves of the antifungal agent were similar to those of blank controls, and the influence of each antifungal agent on the OD was minimal. For the difference in OD between 50% of growth control and the blank control, the curve was the opposite of the blank control, and the OD increased steadily at the wavelengths above 600 nm.

Conclusions: The range between 600 nm and 700 nm was the optimal wavelength for broth microdilution antifungal susceptibility testing, although any wavelength within the visible light spectrum can be used.
KEYWORD
Antifungal susceptibility test, Broth microdilution method, Spectrophotometer, Wavelength