Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0985420070290020217
Laboratory Medicine and Quality Assurance
2007 Volume.29 No. 2 p.217 ~ p.223
The Performance of the Newly Released Automated Chemistry Analyzer Rx Imola
Shin So-Yeon

Kwon Kye-Chul
Park Jong-Woo
Ko Chi-Seon
Koo Sun-Hoe
Abstract
Background: The Rx Imola (Randox, UK) is newly released bench top - fully automated analyzer based on Window XP software with high-throughput (640 tests per hour with ISE) and continuous random access. We evaluated the performance of Rx Imola for the routine chemistry.

Methods: Repeatability (within-day precision), between-day precision, within-device precision, linearity, recovery rates and correlation were evaluated for 19 items including AST, ALT, ALP, GGT, total bilirubin, calcium, phosphorus, albumin, total protein, BUN, creatinine, glucose, amylase, total cholesterol, triglyceride, HDL, LDH, CK and uric acid. Commercialized quality control materials and patient¡¯s sera were used. For correlation study, 747-100 (HITACHI, Japan) and VITROS 950 (Ortho-Clinical Diagnostics, USA) were used as comparative analyzers.

Results: Coefficients of variation (CVs) of all items in repeatability and between-day precision study were below 5%. The linearities were statistically acceptable (R2>0.99) for all items. The recovery rates ranged from 95.7 to 105.3%. The comparison study showed high correlation between Rx Imola and 747-100 or VITROS 950. Correlation coefficients of all items were above 0.99 except HDL and albumin.

Conclusions: This study showed satisfactory results in precision, linearity, recovery rates and comparison studies of Rx Imola. It was expected to be useful for routine chemistry analysis and back up, because of high performance, easy handling and small size.
KEYWORD
Rx Imola, Chemistry analyzer