Laboratory Medicine and Quality Assurance 2008 Volume.30 No. 2 p.273 ~ p.277
|
|
Evaluation of the Tosoh HLC-723 G8 HbA1c Autoanalyzer
|
|
Kwon Ae-Rin
Park Ji-Young Kim Jae-Seok
|
|
Abstract
|
|
|
Background: We evaluated overall performance and analysis time of newly developed HLC-723 G8 (Tosoh corp., Tokyo, Japan) HbA1c autoanalyzerwhich utilizes cation-exchange HPLC method.
Methods: Linearity, precision, correlation with Variant II Turbo (Bio-Rad Laboratories, Hercules, CA, USA), analysis time, labile HbA1c (L-A1c) separation ability and stability of refrigerated sample at 4degrees C were evaluated.
Results: HLC-723 G8 showed good linearity between HbA1c 4.9-12.0% range (R2=0.9995). Within-run and total imprecision (CVs) were less than 1.0%. Correlation with Variant II Turbo and L-A1c separation ability were excellent. Analysis time per sample took 1.0 min. Stability of refrigerated sample at 4degrees C for at least 21 days was good.
Conclusion: HLC-723 G8 showed good analytical performance and its rapid analysis time enables us to support outpatient diagnosis of diabetic patients efficiently.
|
|
KEYWORD
|
|
HbA1c, Ion-exchange HPLC, Tosoh HLC-723 G8
|
|