Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0985420090310010183
Laboratory Medicine and Quality Assurance
2009 Volume.31 No. 1 p.183 ~ p.186
Preparation of Pooled Sera by Using Stirrer
Park Joon-Hong

Kim Yeong-Sic
Abstract
Background: The aim of this study was to compare a new stir method with the conventional refrozen method used to prepare pooled sera.

Methods: We prepared fifteen different pooled sera by using two pooled sera preparation methods. Twenty dispensed sera from each pooled serum were analyzed for glucose, total protein, albumin, aspartate aminotransferase, total bilirubin, triglyceride, blood urea nitrogen, creatinine, sodium, potassium, and chloride concentrations. For stability study, six different pooled sera were stored at room temperature, 4degrees C, and -70degrees C for 3 months and the eleven components were measured for each month.

Results: There was no difference in homogeneity between stir and refrozen pooled sera preparation methods and good stability was observed for -70degrees C storage of all measured components.

Conclusions: The stir method might be more useful than refrozen method to prepare pooled sera.
KEYWORD
Stir, Pooled sera, Quality control
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø