Laboratory Medicine and Quality Assurance 2010 Volume.32 No. 1 p.189 ~ p.196
|
|
A Simple Method to Prepare Anti-Neutrophil Cytoplasmic Antibody Slides
|
|
Won Dong-Il
|
|
Abstract
|
|
|
Background: Neutrophils fixed on a slide are used in indirect immunofluorescence assay (IIF) for anti-neutrophil cytoplasmic antibody (ANCA). A simple method to prepare ANCA slides is described in this report.
Methods: This method uses an overlay method to separate WBC from EDTA whole blood and OctospotTM kits to cytocentrifuge WBC onto 8-well slides. This in-house ANCA IIF was compared with commercial ANCA IIF using 40 sera, and with enzyme linked immunosorbent assay for anti-myeloperoxidase (MPO) antibody (MPO ELISA) using 117 sera.
Results: In ANCA IIF parallel tests using 10 positive and 30 negative sera, in-house slides were in good qualitative agreement, 97.5% (39/40) with commercial kits (¥ê = 0.93). Fluorescence pattern and intensity of positive reaction was also highly concordant between two methods. In MPO ELISA positive cases, the positivity rate of in-house ANCA IIF was 97.8% (45/46), which accounts for its assay sensitivity for the presence of anti-MPO antibody.
Conclusions: In-house ANCA slides was convenient and cost-efficient to be prepared. As these slides were found to have acceptable agreement and sensitivity as compared with commercial kits, in-house ANCA IIF is considered to be useful to screen ANCA in combination with MPO ELISA in routine clinical laboratory practice.
|
|
KEYWORD
|
|
ANCA, Neutrophil, Slide
|
|