Laboratory Medicine and Quality Assurance 2013 Volume.35 No. 2 p.133 ~ p.137
|
|
Artifactual Hyperbilirubinemia due to Paraprotein Interference in a Case of Multiple Myeloma
|
|
Jang Jae-Yun
Kim Han-Jun Uh Young Kim Ju-Won Yoon Kap-Jun Kim Moon-Young Kong Jee-Hyun
|
|
Abstract
|
|
|
When analyzing samples containing paraproteins, various interference effects are encountered in the clinical laboratory. Precipitation of paraproteins mostly interferes with the assays that use photometric detection. Herein, we present a case of a patient with multiple myeloma who had paraproteins and spuriously elevated total bilirubin levels (31.1 mg/dL), which were measured by using Roche total bilirubin assay on the Modular DPE (Roche Diagnostics, Switzerland) chemical analyzer. The total bilirubin concentration reduced from 31.1 mg/dL to 1.5 mg/dL, when tested after three fold dilution of the sample on Modular DPE chemical analyzer.
|
|
KEYWORD
|
|
Hyperbilirubinemia, Paraproteinemias, Positive bias, Chemical analyzer
|
|