Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0985420150370020101
Laboratory Medicine and Quality Assurance
2015 Volume.37 No. 2 p.101 ~ p.109
Evaluation of Analytical Performance of the Cobas 8000 Analyzer Series Module e602
Ko Ki-Woong

Kwon Min-Jung
Woo Hee-Yeon
Park Hyo-Soon
Abstract
Background: The e602, a module of the recently released cobas 8000 modular analyzer series, is an automated system for immunoassays. In this study, we evaluated its analytical performance using 17 immunoassay analytes.

Methods: The Clinical Laboratory Standards Institute guidelines were used to determine the efficiency of the cobas 8000 e602 based on its precision, linearity, assay comparison, and reference range validation. Performance analyses were completed using two levels of quality control materials and pooled sera from our institution. The performance of the cobas 8000 e602 was compared to that of the modular analytics E170. Statistical analyses were performed using Excel 2010 (Microsoft Co., USA) and EP Evaluator Release 10 (Data Innovations, USA).

Results: For all analytes, except level 1 total vitamin D, the coefficients of variation were <5%. The linearity results were within the allowable systemic error limits. The performance comparison revealed that the two systems are comparable, with correlation coefficients (r) >0.975 for all analytes. The reference range validation was also within the allowable criteria.

Conclusions: Taken together, these findings demonstrate that the cobas 8000 e602 analyzer has good precision, linearity, performance comparison, and reference range validation. Thus, e602 is a useful module of the cobas 8000 modular analyzer series.
KEYWORD
Immunoassay, Performance, Evaluation
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø