Laboratory Medicine and Quality Assurance 2016 Volume.38 No. 3 p.137 ~ p.142
|
|
Performance Evaluation of the HM-JACKarc Analyser for Fecal Occult Blood Test
|
|
Park Yu-Mi
Choi Qu-Te Kwon Kye-Chul Koo Sun-Hoe
|
|
Abstract
|
|
|
Background: Fecal occult blood tests have been widely used as a means of gastrointestinal bleeding and colorectal cancer screening. HM-JACKarc (Kyowa Medex Co. Ltd, Japan) is a recently introduced automated fecal occult blood test analyser, which uses latex agglutination method. We evaluated the analytical performance of HM-JACKarc.
Methods: The linearity and precision for HM-JACKarc were evaluated according to the corresponding Clinical and Laboratory Standard Institute guidelines. The comparison study between HM-JACKarc and OC-SENSOR DIANA (Eiken Chemical Co. Ltd., Japan) was done with stool specimens.
Results: The linearity was good (R2=0.999) and the coefficients of variation of within-day precision and between-day precision were 5.2% and 4.9%, respectively, in low concentration and 2.7% each in high concentration. The concordance rate between HM-JACKarc and OC-SENSOR DIANA was 99.0% (198 out of 200).
Conclusions: HM-JACKarc showed excellent performance in linearity, precision, and comparison studies. Therefore, it appears to be a useful automated fecal occult blood test analyser.
|
|