Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0985420170390040178
Laboratory Medicine and Quality Assurance
2017 Volume.39 No. 4 p.178 ~ p.180
Performance Evaluation of the Autokit Total Ketone Bodies
Lee Nu-Ri

Ha Chae-Ok
Yun Hyoung-Joo
Woo Kyung-In
Abstract
In this study, the Autokit Total Ketone Bodies kit (Wako Pure Chemical, Japan), a total ketone measurement assay using an enzymatic method, was evaluated using a Roche Cobas e702 instrument (Roche Diagnostics, Germany). Precision, linearity, carryover, and reference range verification were evaluated with reference to Clinical Laboratory Standards Institute guidelines. Standard materials provided by the manufacturer and patient samples were used for the evaluation. The precision and carryover of the evaluation result satisfied the acceptance criteria. Linearity was also acceptable at more than 0.99. The quantitative Autokit Total Ketone Bodies kit is precise, and can be widely used in clinical laboratories.
KEYWORD
Ketone bodies, 3-hydroxybutyric acid, Evaluation studies as topic
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø