KMID : 0985420190410020075
|
|
Laboratory Medicine and Quality Assurance 2019 Volume.41 No. 2 p.75 ~ p.81
|
|
Annual Report of the Korean Association of External Quality Assessment Service on Urinalysis and Fecal Occult Blood Testing (2018)
|
|
Jeon Chang-Ho
Lee A-Jin Kim Sang-Gyung Suh Hun-Seok Bae Young-Cheol
|
|
Abstract
|
|
|
In 2018, external quality assessment trials for urinalysis and fecal occult blood (FOB) were performed using 1,590 participants. Urine chemistry tests were performed thrice while urine sediment and FOB tests twice. Urine chemistry tests comprised of pH, protein, glucose, ketone body, bilirubin, blood, urobilinogen, nitrite, leukocyte, and specific gravity analyses. The results of urine chemistry and specific gravity tests showed accuracy rates >95%, except for the pH test. The accuracy rate of urine sediments was low, especially for atypical calcium oxalate crystal and red blood cell cast. In the FOB quality test, reagents showed accuracy rates >90%, except for SD and GC Genedia FOB reagents. In the FOB quantitative test, Alfresa NS-Plus C instrument showed falsely high values in the FOB negative specimens.
|
|
KEYWORD
|
|
Quality control, Urinalysis, Occult blood, Accuracy
|
|