Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1001219970030010126
Journal of the Korean Society of Digital Medical Imaging Technology
1997 Volume.3 No. 1 p.126 ~ p.132
Lee Hyung-Jin


Lee Won-Hong
Kim Geon-Joong
Abstract
Purpose: To apply to Program of Auto processor quality control after comparison of Film density variations with amendments to Auto density by using Check density program and Adjust density program of calibration mode into the Laser imager.

Methods: Observe Check and Adjust density variations on the Control chart with standard step and value during seven months from December, 1995 to June, 1956 extending twice a week. (1) Measure density value on the steps after printing out 17-step sensitometric pattern of the Check density program. (2) In the same way, measure density values after amending density by using Adjust density program If they are exceeding allowable error limit.

Results: In case of Check density program, the exceeding limit rates of Density difference(DD) and Middle density(MD) are: FL-IM3543 DD=75%. MD=72.5%, FL-IMD DD=0%. MD=30.8%(14.5%) After amending density by using Adjust density program, the exceeding limit rates of all both Laser imager were zero percent. The standard deviations are show lower FL-IM D than FL-IM3543 on the Check density control chart, but higher on the Adjust density control chart.

Conclusion: (1) Check density variations by printingout sensitometric pattern extending once a week at least for quality control of the Laser imager. (2) In case of a dusty place, check the Laser beam transmission after cleaning Laser optical unit extending once a month. (3) Be sure to measure and check density values by using adjust density program if they are exceeding allowable error limit. (4) Maintain much better film density by performing the adjust density program even if check density values are existed within normal limit.
KEYWORD
Laser imager, Check density and Adjust density
FullTexts / Linksout information
 
Listed journal information