KMID : 1001220090110010015
Journal of the Korean Society of Digital Medical Imaging Technology
2009 Volume.11 No. 1 p.15 ~ p.20
Analysis of Patient Exposure dose with Glass Dosimeter
Kim Jae-In

Choi Weon-Keun
Chang Sung-Won
Oh Chang-Seoph
Lee Kwan-Sup
Ha Dong-Yoon
Abstract
Far reducing medical radiation exposure and managing patient doses, Entrance surface doses(ESDs) were measured at Diagnostic Radiology Department in ASAN medical center, also we determined and compared with the Diagnostic Reference Level(DRL) of some other countries. ESDs were measured far the most common types of X-ray procedures, such as chest PA, lumbar spine AP, lumbar spine lateral, Pelvis AP, Skull PA. ESDs were measured by Glass dosimeter and Unfors Xi meter. Those were applied collimation center of phantom¡¯s entrance skin surface. The results of ESDs were compared Glass dosimeter with Unfors Xi meter. Those were measured within 5% statistical difference. It seemed well agreement at two devices. In most cases ESDs measured far the different types of X ray procedures were found to be lower than the DRL of IAEA, but ESDs on chest PA, lumbar spine AP, lumbar spine lateral, Pelvis AP, Skull PA were proximity ar excesses at DRL of advanced country. Through this study, we need an investigation and improvement at present diagnostic radiology exam system. Also, radiologists make an effort to reduce patient dose and having a technical skill.
KEYWORD
Radiation exposures, Entrance surface dose, Diagnostic reference levels