Journal of the Korean Society of Digital Medical Imaging Technology
2009 Volume.11 No. 2 p.51 ~ p.57
A Study on Comparative Analysis of Diffusion Weighted Image Examination before and after Contrast Injection
Goo Eun-Hoe

Abstract
The purpose of this study would evaluate if having clinical effects on diffusion image with quantitative analysis through ADC values of brain¡¯s normal tissue and lesions before and after contrast injections using a 3.0T. From November in 2007 until December in 2008, a total of 32 patient was performed on 3.0T(Signa Excite, GE Medical System, USA) with the normal or lesions in the patient who requests diffusion weighted image with 8channel head coil. The pulse sequence was used with spin echo EPI(TR: 10000msec, TE: 72.2 msec, Matrix: 128*128, FOV: 240 mm, NEX: 1, diffusion direction: 3, b-value: 1000). Measurement results of ADC values on lesions, CSF, white matter, gray matter, lesions after contrast injection were measured less 75% than before contrast injection, infarction: 100%, CSF: 78%(high), white matter: 71.4%(low), gray matter: 50%(high, low). The results of paired t-test on the deference of ADC values which statically is significant in three(lesions, CSF, white matter)regions except for white matter(p<0.05). Quantitative analysis of lesions, CSF, white matter, gray matter have difference on all regions. ADC values were low in lesions and white matter, normal CSF after contrast injection commonly is high than before contrast injection, ADC values which white matter were high and low (50:50) after contrast injection. 3.0T diffusion weighted image clinically supposed that performing DWI examination after contrast injection was not desirable because of having effects on brain tissue.
KEYWORD
Diffusion weighted image, Apparent diffusion coefficient, T2 weighted image
FullTexts / Linksout information
 
Listed journal information