KMID : 1001220090110020101
Journal of the Korean Society of Digital Medical Imaging Technology
2009 Volume.11 No. 2 p.101 ~ p.107
A Comparison of Dose in Changed Technique Factor Using X-ray Imaging System
Han Dong-Kyoon

Ko Shin-Gwan
Sun Jong-Lyool
Yoon Surk-Hwan
Chung Jae-Eun
Abstract
With the recent development of diagnosis using radiation and increasing demand of the medical treatment, we need to minimize radiation exposure dose. So, This is the method which reduce patient dose by measuring surface dose of radiographic change factor and by comparing theoretical and actual dose, when we take an X-ray which is generally used. By changing the factor of kV, mAs, FSD, whose range is 60 to 120 kV, 20 to 100 mAs, 80 to 180 cm, we compared theoretical surface dose with actual surface dose calculated by the simple calculation program, Bit system, and NDD-M method As a result, when kV and mAs were higher, theoretical surface dose and actual surface dose were more increased. but the higher FSD was, the more decreased surface dose was. According to this, the error were measured about 0.1 to 0.2 mGy in low dose part and about 0.7 to 1.5 mGy in high dose part. Therefore, this shows that theoretical surface dose calculation method is more correct in low dose part than in high dose part. In conclusion, we will have to make constant efforts which can reduce patient and radiographer¡¯s exposure dose, studying methods which can predict patient¡¯s radiation exposure dose more exactly.
KEYWORD
Radiation exposure dose, Bit system, NDD-M method