Abstract
Carried out an experiment for quantitative evaluation about intraoral sensor among oral dental imaging equipments in PACS environment. For evaluation, testing environment & evaluation criteria are established that refer to the ¡¯IEC 62220-1-2¡¯ and experimented with set up the standard radiation penetration that correspond with RQA-3(IEC 61267). Results of experiments using the image J, derived the ESF(edge spread function), LSF (line spread function) and calculated the MTF(modulation transfer function) finally. As a result, the MTF that 0.1, 0.2sec are 10% about 10 lp/mm and 0.32 sec is the 10% about 9 lp/mm in level represents the value of the MTF. Change of scan condition in dental environmental, according to the MTF value taken note that no changes can be seen enough. However, the dimensions of each other size , size intraoral sensors for 3 lp/mm, respectively 40%, 90% of the note might have been the difference between the value of MTF, in accordance with standard sensors might note differences could be observed.
KEYWORD
traoral sensor, Image Evaluation, MTF(modulation transfer function)