Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1001220100120010005
Journal of the Korean Society of Digital Medical Imaging Technology
2010 Volume.12 No. 1 p.5 ~ p.8
Measurement of Skin Dose Distribution for the Mobile X-ray Unit Collimator Shielding Device
Hong Seon-Sook

Kim Deuk-Yong
Abstract
Opened a court in February 10, 2006, a rule of safety management of the diagnosis radiation system was promulgated for safety of the radiation worker, patients and patients¡¯ family members. The purpose of this rule is to minimize the risk of being exposed to radiation during the process of handling X-ray. For this reason, we manufactured shielding device of mobile X-ray unit collimator for diminution of skin dose. Shielding device is made to a thickness of Pb 0.375mm. For portable chest radiography, we measured skin dose 50cm from center ray to 200cm at intervals of 20cm by Unfors Xi detector. As a result, a rule of safety management of the diagnosis radiation system has been strengthened. But there are exceptions, such as ER, OR, ICU to this rule. So shielding device could contribute to protect unnecessary radiation exposure and improve nation¡¯s health.
KEYWORD
Radiation, Mobile X-ray, Shielding device, Collimator
FullTexts / Linksout information
 
Listed journal information