Journal of the Korean Society of Digital Medical Imaging Technology
2010 Volume.12 No. 2 p.97 ~ p.101
Analysis of Space Radiation Dose Rate using portable X-ray Generating Device for Abdomen
Park Chang-Hee

Abstract
This experimental study is carried out one of the General Hospital in Kyungbok providence. Abdomen Phantom being located Anterior-posterior(AP) position on portable bed, and the portable X-ray generating device was placed the phantom at direction. The experiment were set 65 kVp, 10 mAs, , 100 cm(FOD) for the measurement. Digital proportional counting tube survey meter was used for measuring the space scatter dose. Measurement points of horizontal distribution was set up at interval by increasing 50 cm radius of upside, downside, left and right. Vertical distribution of measurement points were set up for the vertical plane with a radius of at intervals with 50cm increments. It is concluded that longer distance from the soure of X-ray significantly decrease radiation dose to the patient and use of the radiation protection device should be applied in clinical practice to reduce dose to the patient.
KEYWORD
Portable X-ray generating device, Radiation dose, Abdomen