Journal of the Korean Society of Digital Medical Imaging Technology
2010 Volume.12 No. 2 p.133 ~ p.138
A Study of Quality Control Environment of Mammography
Hwang In-Sun

Kim Young-Keun
Joo Hyung-Yoon
Abstract
The purpose of this study was to survey and test quality control of mammography system. The conclusion of this study is as follows ; First, The rate of pass for phantom image test shows that Film-Screen mammography system(F/S) and computed mammography system(CR) is 80%, Indirect digital mammography system(DR) is 100%. Second, The test of exposure dose shows that F/S is 921 mR. CR is 1,140 mR, DR is 474 mR. The grade of this testament is CR > F/S > DR. Third, The test of average glandular dose shows that F/S is 1,336 mGy, CR is 1,635 mGy, DR is 1,26 mGy. The grade of this testament is CR > F/S > DR. Fourth, The testament of resolution shows as follows F/S is 11~13 Lp/mm, CR is 4~5 Lp/mm, and DR is 5~7 Lp/mm(F/S > DR > CR) Fifth, The survey of projection, cassette, development and reading shows that user are indifference.
KEYWORD
ACR phantom, Exposure dose, Average glandular dose