KMID : 1001220120140020047
Journal of the Korean Society of Digital Medical Imaging Technology
2012 Volume.14 No. 2 p.47 ~ p.56
Study on dose and image quality by Added filter and Grid change when exam abdominal fluoroscopy
Hong Seon-Sook

Kang Kyeong-Mi
Seong Min-Sook
Lee Jong-Woong
Abstract
Amount of radiation exposure by seeing through fluoroscopy examination while is many patient exposure administration and unprepared misgovernment be. In this study, abdominal fluoroscopy during the scan, the dose and image quality change according to the use of grid and added filter optimized by measuring the test condition is proposed. Uses seeing through fluoroscopy examination equipment of Image Intensifier of Easy Diagnost Eleva (Philips), under tube type and uses Human phantom and measures average area dose according to grid insertion existence and nonexistence and added filter kind change. Measure sum of 29 organ dose and effective dose through PCXMC imagination simulation program and image J program through noise, SNR, image distortion was measured. Resolution, sharpness, and analyzed using the MTF curves. Fluorography the grid to insert the filter thickness and thickening and increased the average area dose and organ doses and effective dose. In the case of spot examination, when inserted grid, average area dose and organ dose and effective dose increased. Filter thickens the average area dose decreased, but the organ doses and effective dose were increased when use 0.2mmCu+1mmAl filter, decreased slightly. Noise and SNR measurements without inserting the gird, if you do not use the added filter was the lowest and when measure the distortion, 0.1mmCu+1mmAl filter was no difference of image quality in case insert grid was judged that when did not use occasion added filter that do not use grid, difference of image quality does not exist. Did not show a big difference, according to the grid and uses of the added filter sharpness, and resolution. Patient dose increases with factors that reduce the quality of the image so reckless grid and the use of the added filter when abdominal fluoroscopy examination should be cautious in using.
KEYWORD
Abdominal fluoroscopy, grid, added filter, dose, image quality
FullTexts / Linksout information
 
Listed journal information