Journal of the Korean Society of Digital Medical Imaging Technology
2013 Volume.15 No. 1 p.21 ~ p.26
Comparative study of dose due to the change of fluoroscopy pulse rate of Epidural Injection treatment time
Seo Jeong-Beom

Oh Dong-Hoon
Lee Jeong-Beom
Lee Jong-Woong
Abstract
To be investigated and measures to reduce the medical exposure of patients to change the Epidural Injection time Fluoroscopy Pulse Rate in this study. Was conducted in 50 patients who underwent surgery Epidural Injection performed by interventional care of Konkuk University Hospital from January to April 2013. The treatment time with the change of Pulse rate, is measured in minutes fluoroscopy time, and measured the area dose (¥ìGym2) and depth dose (mGy). Using the Image J program, to measure the PSNR and SNR. The fluoroscopy time as a result surgery, there was no significance in the statistical analysis, and depth dose is 34.3 to 34.9%, was reduced from 35.8 to 38.7% the area dose. It is possible to reduce the appropriate Pulse rate, to reduce the dose without statistical analysis significance fluoroscopy time.
KEYWORD
Pulse rate, PSNR, DAP, dose