Journal of the Korean Society of Digital Medical Imaging Technology
2013 Volume.15 No. 2 p.39 ~ p.44
Study on Radiation dose in according to Magnification¡¯s rate in fluoroscopy
Kang Kyeong-Mi

Hong Seon-Sook
Seong Min-Sook
Song Woon-Heung
Abstract
Purpose : The purpose of this study is the magnification rates depending on the area of patient dose (DAP) and glass dosimeter see the change of the dose according to the dose characteristics of low-magnification aims to raise standards.

Materials and Method : Direct DR equipment Sonialvision DAR-8000f, Shimadzu was used, the patient entrance dose measurements to the surface of the Rando Phantom of the neck and the abdomen was placed on the Xi unfors. glass dosimeter for measuring organ doses at the same time the Rando Phantom of the major organs in place by inserting a 9 ", 12", 15 ", 17" and 30 seconds for each magnification were measured according in fluoroscopy. DAP meter area of the patient dose was measured.

Result : Esophagography at 17" 143% than 9"magnification the average area dose was increased. Organ dose of Esophagography at 17" was decreased 25.32% than 9" magnification. UGI at 17" was increased 129.73% DAP than 9" magnification. Organ dose of UGI at 17" was decreased 23.32% than 9" magnification. Where the major organs of magnification at 17" were decreased(lung -25.96%, stomach -33.09%, spleen -27.81%, liver -4.92%) than 9" magnification.
Conclusion : Expected to get better quality image While using the proper magnification, and have recognition that difference Organ doses and DAP meter in fluoroscopy.
KEYWORD
Magnification, DAP, Organ dose